REED Instruments R2810-NIST thermocouple calibrator can source & measure voltage and eight different thermocouples (R, S, B, E, K, J, T & N). It is ideal for testing and calibrating temperature sensors in various applications like chemical production, petroleum refineries, manufacturing & power production.
Working Mechanism:
- Insert one end of the test lead into the INPUT terminal of the meter and connect the other end to the OUTPUT terminal of the instrument under test.
- Press the 'INPUT/OUTPUT' button to select the input function.
- Press the FUNC button to choose between thermocouple or DC voltage measurement.
- 'R' and 'Deg. C' will appear on the display.
- For thermocouple measurement, press the RANG button to select the type of thermocouple.
- While measuring, the LCD will show '000.00', followed by the measurement result.
Features:
- REED Instruments R2810-NIST thermocouple calibrator has a 6-digit LCD for easy observations.
- It allows the user to switch temperature units between Celcius and Fahrenheit.
- This unit has a low battery indicator that notifies users about the battery status.
- It has an auto shut-off feature that turns off the calibrator if left idle for more than 15 minutes. This calibrator provides internal cold junction compensation for accurate readings.
- It has a ZERO button to turn the output to 0 degrees C or 0V.
Compatible Accessories:
- REED Instruments CA-05A Soft Carrying Case: This case is ideal for storing and transporting the calibrator. It has a main instrument compartment and a side pouch for accessories.
Standards and Approvals:
Frequently Asked Questions:
Q. What are the ideal operating conditions for this calibrator?
A.
- Temperature - 32 to 122 degrees F
- Relative humidity - 0 to 85%
Q. How do you switch temperature units on this REED Instruments R2810-NIST calibrator?
A.
- Turn Off the meter and press the power button to display the full screen.
- Press and hold the RANG button quickly when the instrument is in the maintenance state (indicated by "1.8.8.888").
- 'AP-XX' will appear on display.
- Press the FUNC button to check for the already set unit (Celcius / Fahrenheit).
- Press the vertical arrow keys to toggle between 'tC-0F' for Fahrenheit and 'tC-0C' for Celsius.
- Press the RJ-ON button to save the required setting.
Q. How to enable / disable auto power-off?
A.
- Turn Off the meter and press the power button to display the full screen.
- Press and hold the RANG button quickly when the instrument is in the maintenance state (indicated by "1.8.8.888").
- 'AP-XX' will appear on display.
- Press the downward arrow key to toggle between 'AP-ON' and 'AP-OF'.
- 'AP-ON' indicates that the automatic power-off is enabled.
- 'AP-OF' indicates that the automatic power-off is disabled.
- Press the RJ-ON button to save the required setting.